Model VD 56 | Vacuum drying chambers for non-flammable solvents
Benefits
- Fast, gentle drying
- Optimum heat transfer through large thermal conducting plates
- Safe work thanks to proven safety concept
- Also option of modular complete system with vacuum pump and pump chamber
Product description
BINDER’s VD series vacuum drying chambers deliver impressive performance through their precise temperature control and gentle drying. The patented expansion rack technology from BINDER also optimizes heat transfer. The shelves can be repositioned for flexible placement and the interior of the chamber is easy to clean.
Important characteristics
- Temperature range: +9°C above ambient temperature to +220 °C
- Controller with digital display of pressure and temperature
- Program-controlled drying monitoring with automatic ventilation at end of process
- Internal data logger, measured values can be read out in open format via USB
- 1 aluminum expansion rack, can be custom-positioned
- Inert gas connection
- Universal access port DN 16 - for the introduction of external measuring devices
- Shatterproof, spring-mounted safety glass panel
- Large viewing window with option of interior lighting
- Computer interface: Ethernet
Technical specifications
Data | ||
Designation | VD056-230V | VD056UL-120V |
Article Number | 9630-0002 | 9630-0006 |
Option model | Standard | Standard |
Performance Data Temperature | ||
Temperature range | +9 °C above ambient temperature to 220 °C | +9 °C above ambient temperature to 220 °C |
Temperature uniformity at 100°C | 1.2 ± K | 1.2 ± K |
Temperature fluctuation at 100°C | 0.1 ± K | 0.1 ± K |
Heating up time to 100°C | 80 min | 80 min |
Vacuum Data | ||
Leckrate | 0,01 bar/h | 0,01 bar/h |
Temperature fluctuation over time at 50 mbar | 1 ± mbar | 1 ± mbar |
Temperature fluctuation over time at 400 mbar | 70 ± mbar | 70 ± mbar |
Electrical data | ||
Rated Voltage | 200...230 V | 100...120 V |
Power frequency | 50/60 Hz | 50/60 Hz |
Nominal power | 1,4 kW | 1,4 kW |
Unit fuse | 8,0 A | 16,0 A |
Phase (Nominal voltage) | 1~ | 1~ |
Connections | ||
Vacuum connection with small flange | 16 DN mm | 16 DN mm |
Measuring access port with small flange | 16 DN mm | 16 DN mm |
Compressed air connection for pressure-encapsulation | 5 NW | 5 NW |
Dimensions and weights | ||
Interior volume | 55 L | 55 L |
Net weight of the unit (empty) | 96 kg | 96 kg |
Load per rack | 20 kg | 20 kg |
Permitted load | 60 kg | 60 kg |
Wall clearance back | 100 mm | 100 mm |
Wall clearance sidewise | 70 mm | 70 mm |
Housing dimensions not incl. fittings and connections | ||
Width net | 638 mm | 638 mm |
Height net | 815 mm | 815 mm |
Depth net | 461 mm | 461 mm |
Internal Dimensions | ||
Interior width | 400 mm | 400 mm |
Interior height | 400 mm | 400 mm |
Interior depth | 343 mm | 343 mm |
Unit doors | 1 | 1 |
Environment-specific data | ||
Sound-pressure level | 40 dB(A) | 40 dB(A) |
Energy consumption at 100°C | 180 Wh/h | 180 Wh/h |
Fixtures | ||
Number of shelves (std./max.) | 1/5 | 1/5 |