Model WIC 2 | Walk-in-chambers for stability testing according to ICH (Q1A)
Benefits
- Stable and uniform environment
- Fully complies with Pharma regulations (ICH)Equipment
- Longevity - corrosion resistant
- Ease of operation
- Environmental friendly refrigerant with GWP value 547
- Maintenance possible with the least possible disruption to the climate conditioning
- Perfectly meets requirements in Pharma industry
- 3 years warranty
Product description
The walk-in-chambers offer a wide range of features and benefits to perfectly meet requirements in stability testing. Perfect test conditions are maintained in a wide range of temperature from 10°C to 50°C and relative humidity of 20% to 90% RH.
Important characteristics
- Relative humidity accuracy of ± 2.5%
- Temperature accuracy of ± 1.5°C
- Duct work for proper distribution of air
- Highly insulated room panels with U value of only 0,21 W/m2K
- LED Lighting with motion detector
- High quality room elements, made in Germany
- Interior completely stainless steel
- Exterior service accessibility
- Room Elements: Galvanized sheet steel, polyester coated, white, similar to RAL 9010, interior surfaces in stainless steel,
4 inch / 100 mm PU foam U value: 0,21W/m2K - Floor: Floor overlay is chrome-nickel steel, pattern-rolled 5WL
- Surface load: 50.000 N/m2
Technical specifications
Data | |||
Designation | WIC2-230 V | WIC2-208V | WIC2C-208V |
Article Number | 9800-0012 | 9800-0003 | 9800-0004 |
Option model | Standard | Standard | with conditioner on ceiling |
Performance Data Temperature | |||
Temperature range | {3299}...{3298} | {3299}...{3298} | {3299}...{3298} |
Performance Data Climate | |||
Temperature range with humidity | 10...50 °C | 10...50 °C | 10...50 °C |
Humidity range | 20...90 % r.F. | 20...90 % r.F. | 20...90 % r.F. |
Temperature uniformity at 25°C and 60% RH | 1.5 ± K | 1.5 ± K | 1.5 ± K |
Temperature uniformity at 40°C and 75% RH | 1.5 ± K | 1.5 ± K | 1.5 ± K |
Temperature fluctuation at 25°C and 60% RH | 0.5 ± K | 0.5 ± K | 0.5 ± K |
Temperature fluctuation at 40°C and 75% RH | 0.5 ± K | 0.5 ± K | 0.5 ± K |
Electrical data | |||
Rated Voltage | 230 V | 208...230 V | 208...230 V |
Power frequency | 50 Hz | 60 Hz | 60 Hz |
Nominal power | 2,7 kW | 2,7 kW | 2,7 kW |
Unit fuse | 20,0 A | 20,0 A | 20,0 A |
Phase (Nominal voltage) | 1~ | 1~ | 1~ |
Dimensions and weights | |||
Interior volume | 18,523 L | 18,523 L | 18,523 L |
Net weight of the unit (empty) | 981 kg | 981 kg | 981 kg |
Housing dimensions not incl. fittings and connections | |||
Width net | 2,600 mm | 2,600 mm | 2,600 mm |
Height net | 2,470 mm | 2,470 mm | 2,470 mm |
Depth net | 3,600 mm | 3,600 mm | 3,600 mm |
Internal Dimensions | |||
Interior width | 2,400 mm | 2,400 mm | 2,400 mm |
Interior height | 2,270 mm | 2,270 mm | 2,270 mm |
Interior depth | 3,400 mm | 3,400 mm | 3,400 mm |
Unit doors | 1 | 1 | 1 |
Environment-specific data | |||
Energy consumption at 40°C and 75% RH | 1,140 Wh/h | 1,140 Wh/h | 1,140 Wh/h |