Model ED 23 | Drying and heating chambers with gravity convection
Benefits
- Uniform drying conditions thanks to APT.line™ technology
- Identical test conditions throughout the chamber interior independent of sample size and quantity
- Outstanding thermal insulation saves operation costs
Product description
The strengths of a BINDER ED series drying chamber include routine drying and sterilization tasks up to 300 °C. Thanks to the gravity convection, all thermal processes which use this drying chamber are highly efficient. The ED series ensures fast, uniform drying.
Important characteristics
- Temperature range: room temperature plus 5 °C to 300 °C
- APT.line™ preheating chamber technology
- Gravity convection
- Adjustable exhaust air flap
- Controller with timer functions
- 2 chrome-plated racks
- Class 2 independent adjustable temperature safety device (DIN 12880) with visual alarm
Technical specifications
Data | ||||
Designation | ED023-230V | ED023-230V-I | ED023UL-120V | ED023UL-120V-I |
Article Number | 9010-0190 | 9010-0191 | 9010-0192 | 9010-0193 |
Option model | Standard | with RS 422 | Standard | with RS 422 |
Performance Data Temperature | ||||
Temperature range | +5 °C above ambient temperature to 300 °C | +5 °C above ambient temperature to 300 °C | +5 °C above ambient temperature to 300 °C | +5 °C above ambient temperature to 300 °C |
Temperature uniformity at 150°C | 2.8 ± K | 2.8 ± K | 2.8 ± K | 2.8 ± K |
Temperature fluctuation at 150°C | 0.5 ± K | 0.5 ± K | 0.5 ± K | 0.5 ± K |
Heating up time to 150°C | 28 min | 28 min | 28 min | 28 min |
Recovery time after door was opened for 30 s at 150°C | 5 min | 5 min | 5 min | 5 min |
Air change data | ||||
Air change rate at 150°C | 13 x/h | 13 x/h | 13 x/h | 13 x/h |
Electrical data | ||||
Rated Voltage | 230 V | 120 V | ||
Power frequency | 50/60 Hz | 50/60 Hz | 60 Hz | 60 Hz |
Nominal power | 0,8 kW | 0,8 kW | 0,8 kW | 0,8 kW |
Unit fuse | 10,0 A | 10,0 A | 12,5 A | 12,5 A |
Phase (Nominal voltage) | 1~ | 1~ | 1~ | 1~ |
Dimensions and weights | ||||
Interior volume | 20 L | 20 L | 20 L | 20 L |
Net weight of the unit (empty) | 27 kg | 27 kg | 27 kg | 27 kg |
Load per rack | 12 kg | 12 kg | 12 kg | 12 kg |
Permitted load | 25 kg | 25 kg | 25 kg | 25 kg |
Wall clearance back | 100 mm | 100 mm | 100 mm | 100 mm |
Wall clearance sidewise | 50 mm | 50 mm | 50 mm | 50 mm |
Housing dimensions not incl. fittings and connections | ||||
Width net | 435 mm | 435 mm | 435 mm | 435 mm |
Height net | 495 mm | 495 mm | 495 mm | 495 mm |
Depth net | 520 mm | 520 mm | 520 mm | 520 mm |
Internal Dimensions | ||||
Interior width | 222 mm | 222 mm | 222 mm | 222 mm |
Interior height | 330 mm | 330 mm | 330 mm | 330 mm |
Interior depth | 300 mm | 300 mm | 300 mm | 300 mm |
Unit doors | 1 | 1 | 1 | 1 |
Environment-specific data | ||||
Energy consumption at 150°C | 148 Wh/h | 148 Wh/h | 148 Wh/h | 148 Wh/h |
Fixtures | ||||
Number of shelves (std./max.) | 2/4 | 2/4 | 2/4 | 2/4 |