Model MKT 720 | Dynamic climate chambers for rapid temperature changes with extended low temperature range
Benefits
- Homogeneous climate conditions thanks to APT.line™ technology
- Comprehensive programming and data acquisition
- Large heated viewing window
Product description
Temperature ranges between -70 °C and 180 °C, with the added benefit of natural simulation, are what make the BINDER MKT series so unique. At the same time, this environmental simulation chamber fulfills the very highest precision and performance requirements for cyclical temperature tests.
Important characteristics
- Temperature range: -70 °C to +180 °C
- 4 zero-voltage relay contacts
- APT.line™ preheating chamber technology
- Programmable condensation protection for test material
- Heated viewing window with LED interior lighting
- BINDER Multi Management Software APT-COM™ Basic Edition
- Troubleshooting system with visual and audible alarms
- Intuitive touchscreen controller with time-segment and real-time programming
- Internal data logger, measured values can be read out in open format via USB
- 2 access ports with silicone plugs: 80 mm, left and right
- Class 2 independent adjustable temperature safety device (DIN 12880) with visual alarm
- 4 castors, two with brakes
- Computer interface: Ethernet
- Adjustable ramp function
- Integrated chart recorder
- Real-time clock
- Door heating
- 1 stainless steel rack
- Inner chamber made of stainless steel
- CFC-free refrigerants R-452A and R-23
- Cooling with cascade compressor cooling unit
Technical specifications
Data | ||
Designation | MKT720-400V | MKT720-480V-C |
Article Number | 9020-0387 | 9020-0365 |
Option model | Standard | with voltage and frequency converter |
Performance Data Temperature | ||
Temperature range | -70...180 °C | -70...180 °C |
Temperature uniformity dependent on set value | 0.1...2.0 ± K | 0.1...2.0 ± K |
Temperature fluctuation dependent on set value | 0.1...0.5 ± K | 0.1...0.5 ± K |
Average heating-up rate according to IEC 60068-3-5 | 4.5 K/min | 4.5 K/min |
Cooling down time from 180°C to -70°C | 120 min | 120 min |
Average cooling down time according to IEC 60068-3-5 | 4.0 K/min | 4.0 K/min |
Electrical data | ||
Rated Voltage | 400 V | 480 V |
Power frequency | 50 Hz | 60 Hz |
Nominal power | 13 kW | 13 kW |
Unit fuse | 32,0 A | 32,0 A |
Phase (Nominal voltage) | 3~ | 3~ |
Dimensions and weights | ||
Interior volume | 734 L | 734 L |
Net weight of the unit (empty) | 610 kg | 610 kg |
Load per rack | 40 kg | 40 kg |
Permitted load | 160 kg | 160 kg |
Wall clearance back | 300 mm | 300 mm |
Wall clearance sidewise | 200 mm | 200 mm |
Housing dimensions not incl. fittings and connections | ||
Width net | 1,580 mm | 1,580 mm |
Height net | 2,005 mm | 2,005 mm |
Depth net | 1,140 mm | 1,140 mm |
Internal Dimensions | ||
Interior width | 1,200 mm | 1,200 mm |
Interior height | 1,020 mm | 1,020 mm |
Interior depth | 600 mm | 600 mm |
Unit doors | 1 | 1 |
Environment-specific data | ||
Sound-pressure level | 65 dB(A) | 67 dB(A) |
Energy consumption at 20°C | 1,900 Wh/h | 1,900 Wh/h |
Fixtures | ||
Number of shelves (std./max.) | 1/11 | 1/11 |