Model 56
Model 56
Dimensions
Heating up rate
Cooling down rate
Heat compensation chart
Product image
Product image
Product image
Product image
Product image
Product image

Model MK 56 | Dynamic climate chambers for rapid temperature changes

Benefits
  • Homogeneous climate conditions thanks to APT.line™ technology
  • Comprehensive programming and data acquisition
  • Large heated viewing window

Configure your product with options and accessories and request it directly here.

Product description

The BINDER MK series chamber is suitable for all heat and cold tests between -40 °C and 180 °C. The MK is a clever alternative to complex individual solutions for cyclical temperature tests.

Important characteristics

  • Temperature range: -40 °C to +180 °C
  • APT.line™ preheating chamber technology
  • Programmable condensation protection for test material
  • Heated viewing window with LED interior lighting
  • BINDER Multi Management Software APT-COM™ Basic Edition
  • Troubleshooting system with visual and audible alarms
  • Intuitive touchscreen controller with time-segment and real-time programming
  • Internal data logger, measured values can be read out in open format via USB
  • Unit self-test for comprehensive status analysis
  • Access port with silicone plug, 50 mm, left
  • Class 2 independent adjustable temperature safety device (DIN 12880) with visual alarm
  • 4 castors, two with brakes
  • Computer interface: Ethernet
  • Adjustable ramp function
  • Integrated chart recorder
  • Real-time clock
  • 1 stainless steel rack
  • Inner chamber made of stainless steel
  • CFC-free refrigerant R-452A
  • Cooling with compressor cooling unit

Technical specifications

Data
Designation

MK056-230V

MK056-240V

Article Number9020-03749020-0388
Option model

Standard

Standard

Performance Data Temperature
Temperature range-40...180 °C-40...180 °C
Temperature uniformity dependent on set value0.5...1.5 ± K0.5...1.5 ± K
Temperature fluctuation dependent on set value0.1...0.5 ± K0.1...0.5 ± K
Average heating-up rate according to IEC 60068-3-55.0 K/min5.0 K/min
Cooling down time from 180°C to -40°C90 min90 min
Average cooling down time according to IEC 60068-3-55.0 K/min5.0 K/min
Electrical data
Rated Voltage230 V240 V
Power frequency50 Hz60 Hz
Nominal power2,8 kW2,8 kW
Unit fuse16,0 A16,0 A
Phase (Nominal voltage)1~1~
Dimensions and weights
Interior volume60 L60 L
Net weight of the unit (empty)165 kg165 kg
Load per rack15 kg15 kg
Permitted load60 kg60 kg
Wall clearance back300 mm300 mm
Wall clearance sidewise200 mm200 mm
Housing dimensions not incl. fittings and connections
Width net720 mm720 mm
Height net1,445 mm1,445 mm
Depth net778 mm778 mm
Internal Dimensions
Interior width400 mm400 mm
Interior height420 mm420 mm
Interior depth350 mm350 mm
Unit doors11
Environment-specific data
Sound-pressure level59 dB(A)59 dB(A)
Energy consumption at 20°C600 Wh/h600 Wh/h
Fixtures
Number of shelves (std./max.)1/41/4

Downloads

Dynamic climate chambers with dry-air purge

For the MK, MKF, MKT, and MKFT series

pdf, 150.06 KB

Constant climate chambers (KBF LQC series) - Top-quality dairy products

Arla Foods Deutschland GmbH

pdf, 802.53 KB

MK | MKT series

pdf, 11.25 MB

Data Sheet Model MK 56

pdf, 589.16 KB

Suggested products

Compare products:

Or add another product.