Model MK 400 | Dynamic climate chambers for rapid temperature changes
Benefits
- Homogeneous climate conditions thanks to APT.line™ technology
- Comprehensive programming and data acquisition
- Large heated viewing window
Product description
The BINDER MK series chamber is suitable for all heat and cold tests between -40 °C and 180 °C. The MK is a clever alternative to complex individual solutions for cyclical temperature tests.
Important characteristics
- Temperature range: -40 °C to +180 °C
- APT.line™ preheating chamber technology
- Programmable condensation protection for test material
- Heated viewing window with LED interior lighting
- BINDER Multi Management Software APT-COM™ Basic Edition
- Troubleshooting system with visual and audible alarms
- Intuitive touchscreen controller with time-segment and real-time programming
- Internal data logger, measured values can be read out in open format via USB
- Unit self-test for comprehensive status analysis
- Access port with silicone plug, 50 mm, left
- Class 2 independent adjustable temperature safety device (DIN 12880) with visual alarm
- 4 castors, two with brakes
- Computer interface: Ethernet
- Adjustable ramp function
- Integrated chart recorder
- Real-time clock
- 1 stainless steel rack
- Inner chamber made of stainless steel
- CFC-free refrigerant R-452A
- Cooling with compressor cooling unit
Technical specifications
Data | ||
Designation | MK400-400V | MK400-480V-C |
Article Number | 9020-0406 | 9020-0445 |
Option model | Standard | with voltage and frequency converter |
Performance Data Temperature | ||
Temperature range | -40...180 °C | -40...180 °C |
Temperature uniformity dependent on set value | 0.1...1.2 ± K | 0.1...1.2 ± K |
Temperature fluctuation dependent on set value | 0.1...0.5 ± K | 0.1...0.5 ± K |
Average heating-up rate according to IEC 60068-3-5 | 5.0 K/min | 5.0 K/min |
Cooling down time from 180°C to -40°C | 115 min | 115 min |
Average cooling down time according to IEC 60068-3-5 | 5.0 K/min | 5.0 K/min |
Electrical data | ||
Rated Voltage | 400 V | 480 V |
Power frequency | 50 Hz | 60 Hz |
Nominal power | 7 kW | 7 kW |
Unit fuse | 16,0 A | 16,0 A |
Phase (Nominal voltage) | 3~/N/PE | 3~ |
Dimensions and weights | ||
Interior volume | 417 L | 417 L |
Net weight of the unit (empty) | 413 kg | 413 kg |
Load per rack | 30 kg | 30 kg |
Permitted load | 150 kg | 150 kg |
Wall clearance back | 500 mm | 500 mm |
Wall clearance sidewise | 300 mm | 300 mm |
Housing dimensions not incl. fittings and connections | ||
Width net | 1,115 mm | 1,115 mm |
Height net | 1,710 mm | 1,710 mm |
Depth net | 1,400 mm | 1,400 mm |
Internal Dimensions | ||
Interior width | 735 mm | 735 mm |
Interior height | 700 mm | 700 mm |
Interior depth | 810 mm | 810 mm |
Unit doors | 1 | 1 |
Environment-specific data | ||
Sound-pressure level | 65 dB(A) | 65 dB(A) |
Energy consumption at 20°C | 1,400 Wh/h | 1,400 Wh/h |
Fixtures | ||
Number of shelves (std./max.) | 1/6 | 1/6 |