Model KB ECO 720 | Cooling incubators with environmentally friendly thermoelectric cooling
Benefits
- Safe: Unique condensate and dry-out protection for all operating conditions
- Reliable: Fail-safe operation due to intelligent thermoelectric tempering
- Smart: Temperatures programmable for intermitted incubation and refrigeration
- Economical: Minimum energy consumption thanks to optimized thermoelectric cooling
Product description
Thanks to thermoelectric cooling technology with patented heat dissipation, the new KB ECO cooling incubator is one of the most energy-efficient cooling incubator on the market. The thermoelectric cooling technology means the KB ECO series is also very quiet during operation.
Important characteristics
- Temperature range: 0°C through +70°C (max. 26°C below the ambient temperature)
- APT.line™ preheating chamber technology
- Electric temperature control thanks to thermoelectric cooling module with innovative heat dissipation
- Inner door made of tempered safety glass
- Inner chamber made completely of stainless steel
- 2 stainless steel racks
- Adjustable condensate protection
- Access port with silicone plug Ø 30 mm
- LCD to display temperature along with additional information and alarms
- Internal data logger, measured values can be read out in open format via USB
- Unit self-test for comprehensive status analysis
- Computer interface: Ethernet
Technical specifications
Data | ||
Designation | KBECO720-230V | KBECO720UL-120V |
Article Number | 9020-0424 | 9020-0427 |
Option model | Standard | Standard |
Performance Data Temperature | ||
Temperature range | (max. 26 °C under ambient temperature) 0...70 °C | (max. 26 °C under ambient temperature) 0...70 °C |
Temperature uniformity at 37°C | 0.1 ± K | 0.1 ± K |
Temperature fluctuation at 37°C | 0.1 ± K | 0.1 ± K |
Max. heat compensation at 40°C | 500 W | 500 W |
Recovery time after door was opened for 30 s at 37°C | 4 min | 4 min |
Electrical data | ||
Rated Voltage | 200...230 V | 100...120 V |
Power frequency | 50/60 Hz | 50/60 Hz |
Nominal power | 1,3 kW | 1,3 kW |
Unit fuse | 16,0 A | 16,0 A |
Phase (Nominal voltage) | 1~ | 1~ |
Dimensions and weights | ||
Interior volume | 700 L | 700 L |
Net weight of the unit (empty) | 272 kg | 272 kg |
Load per rack | 45 kg | 45 kg |
Permitted load | 150 kg | 150 kg |
Wall clearance back | 100 mm | 100 mm |
Wall clearance sidewise | 100 mm | 100 mm |
Housing dimensions not incl. fittings and connections | ||
Width net | 1,250 mm | 1,250 mm |
Height net | 1,925 mm | 1,925 mm |
Depth net | 885 mm | 885 mm |
Internal Dimensions | ||
Interior width | 973 mm | 973 mm |
Interior height | 1,250 mm | 1,250 mm |
Interior depth | 576 mm | 576 mm |
Inner doors | 2 | 2 |
Unit doors | 2 | 2 |
Environment-specific data | ||
Sound-pressure level | 47 dB(A) | 47 dB(A) |
Energy consumption at 37°C | 105 Wh/h | 105 Wh/h |
Fixtures | ||
Number of shelves (std./max.) | 2/15 | 2/15 |