Model KBF-S 720 | Constant climate chambers with large temperature / humidity range
Benefits
- Safe thanks to climatic homogeneity that far exceeds the accuracy required by ICH guidelines, even with a fully loaded unit.
- Reliable thanks to failsafe operation without compromise. The interior and Longlife evaporator plate are made entirely from stainless steel.
- Smart, as a wide range of accessories makes it highly compatible for adaptation to specific customer requirements.
- Economical thanks to maximum effective volume which allows for 30% larger load compared to the competition.
Product description
The Solid.Line KBF-S climate chamber is a custom-made solution for long-term testing and accelerated stability testing of pharmaceutical products in accordance with the ICH Q1A guideline. The corrosion-free and condensation-free interior also ensures exceptionally stable temperature and moisture levels, even when fully loaded.
Important characteristics
- Temperature range: 0 °C to +70 °C
- Humidity range: 20% to 80% RH
- APT.line™ preheating chamber technology
- Humidity regulation with capacitive humidity sensor and steam humidification
- LCD to display temperature and humidity along with additional information and alarms
- Internal data logger, measured values can be read out in open format via USB
- Unit self-test for comprehensive status analysis
- Inner chamber made completely of stainless steel
- 2 stainless steel racks, incl. shelf supports
- Access port with silicone plug, 30 mm, left
- Robust twin casters with brakes, from 240 L
- Class 3.1 independent temperature safety device (DIN 12880) with visual and audible temperature alarm
- Computer interface: Ethernet
- Independent water supply via tank
Technical specifications
Data | ||
Designation | KBFS720-230V | KBFS720UL-240V |
Article Number | 9020-0368 | 9020-0369 |
Option model | Standard | Standard |
Performance Data Temperature | ||
Temperature range | 0...70 °C | 0...70 °C |
Max. heat compensation at 40°C | 550 W | 550 W |
Performance Data Climate | ||
Temperature range with humidity | 10...70 °C | 10...70 °C |
Humidity range | 20...80 % RH | 20...80 % RH |
Temperature uniformity at 25°C and 60% RH | 0.3 ± K | 0.3 ± K |
Temperature uniformity at 40°C and 75% RH | 0.3 ± K | 0.3 ± K |
Temperature fluctuation at 25°C and 60% RH | 0.1 ± K | 0.1 ± K |
Temperature fluctuation at 40°C and 75% RH | 0.1 ± K | 0.1 ± K |
Recovery time after door was opened for 30 s at 25°C and 60% RH | 16 min | 16 min |
Recovery time after door was opened for 30 s at 40°C and 75% RH | 17 min | 17 min |
Electrical data | ||
Rated Voltage | 200...230 V | 200...240 V |
Power frequency | 50/60 Hz | 50/60 Hz |
Nominal power | 2 kW | 2 kW |
Unit fuse | 16,0 A | 16,0 A |
Phase (Nominal voltage) | 1~ | 1~ |
Dimensions and weights | ||
Interior volume | 700 L | 700 L |
Net weight of the unit (empty) | 284 kg | 284 kg |
Load per rack | 45 kg | 45 kg |
Permitted load | 150 kg | 150 kg |
Wall clearance back | 100 mm | 100 mm |
Wall clearance sidewise | 300 mm | 300 mm |
Housing dimensions not incl. fittings and connections | ||
Width net | 1,250 mm | 1,250 mm |
Height net | 1,925 mm | 1,925 mm |
Depth net | 890 mm | 890 mm |
Internal Dimensions | ||
Interior width | 973 mm | 973 mm |
Interior height | 1,250 mm | 1,250 mm |
Interior depth | 576 mm | 576 mm |
Unit doors | 2 | 2 |
Environment-specific data | ||
Sound-pressure level | 53 dB(A) | 53 dB(A) |
Energy consumption at 40°C and 75% RH | 500 Wh/h | 500 Wh/h |
Fixtures | ||
Number of shelves (std./max.) | 2/15 | 2/15 |