Benefits
- Safe thanks to climatic homogeneity that far exceeds the accuracy required by ICH guidelines, even with a fully loaded unit. The light spectrum is 100% compliant with ICH requirements.
- Reliable thanks to failsafe operation without compromise. The interior and Longlife evaporator plate are made entirely from stainless steel.
- Smart, as a wide range of accessories makes it highly compatible for adaptation to specific customer requirements.
- Economical thanks to maximum effective volume which allows for 30% larger load compared to the competition.
Product description
The main advantage of the BINDER KBF LQC series is the patented light photometry for photostability tests. Two 3D spherical sensors control the light dosage of UV-A and visible light. Once the desired light dosage is achieved, the chamber switches off automatically.
Important characteristics
- Temperature range: 0 °C to +70 °C
- Temperature range with light: +10 °C to +60 °C
- Humidity range: 10 % to 80 % RH
- 3 positionable illumination cassettes with ICH-compliant UV/Vis light source
- Independent light-dose control of UV-A and visible light with spherical sensors
- APT.line™ preheating chamber technology
- Humidity regulation with capacitive humidity sensor and vapor humidification
- Inner chamber made of stainless steel
- BINDER Multi Management Software APT-COM™ Basic Edition
- Intuitive touchscreen controller with time-segment and real-time programming
- Internal data logger, measured values can be read out in open format via USB
- Unit self-test for comprehensive status analysis
- Tight-sealing inner door made of safety glass (ESG) with special TIMELESS coating
- 3 stainless steel racks
- Access port with silicone plug, 30 mm, left
- 4 stable castors, two with brakes
- Class 3.1 independent temperature safety device (DIN 12880) with visual and audible temperature alarm
- Computer interface: Ethernet
- Door heating
Technical specifications
Data | ||
Designation | KBFLQC720-230V | KBFLQC720UL-240V |
Article Number | 9020-0334 | 9020-0335 |
Option model | Standard | Standard |
Performance Data Temperature | ||
Temperature range | 0...70 °C | 0...70 °C |
Temperature range without illumination cassettes | 0...70 °C | 0...70 °C |
Temperature range with 100% illumination | 10...60 °C | 10...60 °C |
Max. heat compensation at 40°C with illumination | 1,000 W | 1,000 W |
Performance Data Climate | ||
Temperature range with humidity and without illumination cassettes | 10...70 °C | 10...70 °C |
Temperature range with humidity and 100% illumination | 20...60 °C | 20...60 °C |
Humidity range without illumination cassettes | 10...70 % RH | 10...70 % RH |
Humidity range with 100% illumination | ...60 % RH | ...60 % RH |
Temperature fluctuation with illumination at 25°C and 60% RH | 0.2 ± K | 0.2 ± K |
Humidity fluctuation with illumination at 25°C and 60% RH | 2 % RH | 2 % RH |
Recovery time after door was opened for 30 s with illumination at 25°C and 60% RH | 1 min | 1 min |
Recovery time after door was opened for 30 s with illumination at 40°C and 75% RH | 5 min | 5 min |
Lightdata per Illumination Cassettes | ||
ICH compliant illumination for photo stability testing | 9000 lx | 9000 lx |
ICH compliant illumination for photo stability testing | 1,5 W/m² | 1,5 W/m² |
Electrical data | ||
Rated Voltage | 200...230 V | 200...240 V |
Power frequency | 50 Hz | 50/60 Hz |
Nominal power | 3,5 kW | 3,5 kW |
Unit fuse | 16,0 A | 16,0 A |
Phase (Nominal voltage) | 1~ | 1~ |
Dimensions and weights | ||
Interior volume | 700 L | 700 L |
Net weight of the unit (empty) | 374 kg | 374 kg |
Load per rack | 45 kg | 45 kg |
Permitted load | 150 kg | 150 kg |
Wall clearance back | 100 mm | 100 mm |
Wall clearance sidewise | 200 mm | 200 mm |
Housing dimensions not incl. fittings and connections | ||
Width net | 1,250 mm | 1,250 mm |
Height net | 1,925 mm | 1,925 mm |
Depth net | 890 mm | 890 mm |
Internal Dimensions | ||
Interior width | 973 mm | 973 mm |
Interior height | 1,250 mm | 1,250 mm |
Interior depth | 576 mm | 576 mm |
Inner doors | 2 | 2 |
Unit doors | 2 | 2 |
Environment-specific data | ||
Sound-pressure level | 53 dB(A) | 53 dB(A) |
Energy consumption at 40°C and 75% RH | 2,350 Wh/h | 2,350 Wh/h |
Fixtures | ||
Number of shelves (std./max.) | 3/12 | 3/12 |
Number of illumination cassettes (std./max.) | 3/3 | 3/3 |