Model MK 115 | Dynamic climate chambers for rapid temperature changes

Benefits
  • Homogeneous climate conditions thanks to APT.line™ technology
  • Comprehensive programming and data acquisition
  • Large heated viewing window
with voltage and frequency converter Please select Model
  • Please select Model
  • with voltage and frequency converter
  • Standard
480 V Please select Voltage
  • Please select Voltage
  • 400 V
  • 480 V

Configure your product with options and accessories and request it directly here.

Product description

The BINDER MK series chamber is suitable for all heat and cold tests between -40 °C and 180 °C. The MK is a clever alternative to complex individual solutions for cyclical temperature tests.

Important characteristics

  • Temperature range:

    -40 °C to +180 °C

  • APT.line™ preheating chamber technology
  • Programmable condensation protection for test material
  • Heated viewing window with LED interior lighting
  • BINDER Multi Management Software APT-COM™ Basic Edition
  • Troubleshooting system with visual and audible alarms
  • Intuitive touchscreen controller with time-segment and real-time programming
  • Internal data logger, measured values can be read out in open format via USB
  • Access port with silicone plug, 50 mm, left
  • Class 2 independent adjustable temperature safety device (DIN 12880) with visual alarm
  • 4 castors, two with brakes
  • Computer interface:

    Ethernet

  • 230 V power socket on the right-side control panel
  • Adjustable ramp function
  • Integrated chart recorder
  • Real-time clock
  • Door heating
  • 1 stainless steel rack
  • Inner chamber made of stainless steel
  • CFC-free refrigerant R-452A
  • Cooling with compressor cooling unit

Technical specifications

DesignationMK115-400VMK115-480V-C
Article Number9020-03759020-0303
Option model

Standard

with voltage and frequency converter

DesignationMK115-400VMK115-480V-C
Temperature range-40...180 °C-40...180 °C
Temperature uniformity dependent on set value0.1...2.0 ± K0.1...2.0 ± K
Temperature fluctuation dependent on set value0.1...0.5 ± K0.1...0.5 ± K
Average heating-up rate according to IEC 60068-3-55.3 K/min5.3 K/min
Cooling down time from 180°C to -40°C90 min90 min
Average cooling down time according to IEC 60068-3-55.0 K/min5.0 K/min

DesignationMK115-400VMK115-480V-C
Rated Voltage400 V480 V
Power frequency50 Hz60 Hz
Nominal power3,5 kW3,5 kW
Unit fuse16,0 A16,0 A
Phase (Nominal voltage)3~/N/PE3~

DesignationMK115-400VMK115-480V-C
Interior volume115 L115 L
Net weight of the unit (empty)255 kg255 kg
Load per rack30 kg30 kg
Permitted load60 kg60 kg
Wall clearance back300 mm300 mm
Wall clearance sidewise200 mm200 mm

DesignationMK115-400VMK115-480V-C
Width net980 mm980 mm
Height net1,725 mm1,725 mm
Depth net865 mm865 mm

DesignationMK115-400VMK115-480V-C
Interior width600 mm600 mm
Interior height480 mm480 mm
Interior depth400 mm400 mm
Unit doors11

DesignationMK115-400VMK115-480V-C
Sound-pressure level62 dB(A)67 dB(A)
Energy consumption at 20°C650 Wh/h650 Wh/h

DesignationMK115-400VMK115-480V-C
Number of shelves (std./max.)1/41/4

Downloads

Dynamic climate chambers with dry-air purge

For the MK, MKF, MKT, and MKFT series

pdf, 150.06 KB

Constant climate chambers (KBF LQC series) - Top-quality dairy products

Arla Foods Deutschland GmbH

pdf, 802.53 KB

MK | MKT series

pdf, 15.88 MB

Data Sheet Model MK 115

pdf, 666.37 KB

Suggested products

Compare products:

Or add another product.