Model LIT MK 720 | Battery test chambers with safety equipment for rapid temperature changes
Benefits
- Safety thanks to extensive protective measures
- Homogeneous climate conditions thanks to APT.line™ preheating chamber
- Extensive programming and data collection
- Excellent overview of the test object thanks to large heated viewing window
Product description
The BINDER MK series chamber is suitable for all heat and cold tests between -40 °C and 180 °C. The MK is a clever alternative to complex individual solutions for cyclical temperature tests.
Important characteristics
- Temperature range: -40°C to +110°C
- Class 2 independent temperature safety device when temperature is set to +120°C
- Reversible stainless steel pressure relief flap installed in the middle on top of the unit
- Enhanced door-locking mechanism with reinforced brackets
- Monitoring of CO, H2, O2, and temperature in testing area atmosphere
- Inertization equipment
- CO2 fire suppression
- APT.line™ preheating chamber technology
- Programmable condensation protection for samples
- Heated viewing window with LED interior lighting
- BINDER Multi Management Software APT-COM™ basic edition
- Intuitive touchscreen controller with time-segment and real-time programming
- Internal data logger, measured values can be read out in open format via USB
- 2 access ports with silicone plugs, 80 mm, left and right
- Class 2 independent adjustable temperature safety device (DIN 12880) with visual alarm
- Computer interface: Ethernet
- 1 stainless steel rack, can be freely positioned
- CFC-free refrigerant R-452A
- Cooling with compressor cooling unit
- TÜV SÜD confirms that the LIT MK 240 and LIT MK 720 chambers are suitable for EUCAR level 6
Technical specifications
Data | ||
Designation | LITMK720-400V | LITMK720-480V-C |
Article Number | 9020-0403 | 9020-0405 |
Option model | Standard | Standard |
Performance Data Temperature | ||
Temperature range | -40...110 °C | -40...110 °C |
Temperature uniformity dependent on set value | 0.3...2.0 ± K | 0.3...2.0 ± K |
Temperature fluctuation dependent on set value | 0.1...0.5 ± K | 0.1...0.5 ± K |
Average heating-up rate according to IEC 60068-3-5 | 4.0 K/min | 4.0 K/min |
Cooling down time from 110°C to -40°C | 100 min | 100 min |
Average cooling down time according to IEC 60068-3-5 | 3.4 K/min | 3.4 K/min |
Electrical data | ||
Rated Voltage | 400 V | 480 V |
Power frequency | 50 Hz | 60 Hz |
Nominal power | 8,7 kW | 8,7 kW |
Unit fuse | 16,0 A | 16,0 A |
Phase (Nominal voltage) | 3~ | 3~ |
Dimensions and weights | ||
Interior volume | 734 L | 734 L |
Net weight of the unit (empty) | 584 kg | 584 kg |
Load per rack | 40 kg | 40 kg |
Permitted load | 160 kg | 160 kg |
Wall clearance back | 300 mm | 300 mm |
Wall clearance sidewise | 200 mm | 200 mm |
Housing dimensions not incl. fittings and connections | ||
Width net | 1,794 mm | 1,794 mm |
Height net | 2,005 mm | 2,005 mm |
Depth net | 1,186 mm | 1,186 mm |
Internal Dimensions | ||
Interior width | 1,200 mm | 1,200 mm |
Interior height | 1,020 mm | 1,020 mm |
Interior depth | 600 mm | 600 mm |
Unit doors | 1 | 1 |
Environment-specific data | ||
Sound-pressure level | 65 dB(A) | 67 dB(A) |
Energy consumption at 20°C | 1,900 Wh/h | 1,900 Wh/h |
Fixtures | ||
Number of shelves (std./max.) | 1/11 | 1/11 |